The sensitivity of the measurement is affected by the location of the
measured area of the sample relative to the camera centre line, marked
‘C’ on Figure 1. The reflected light is more sensitive to small variations
in the surface topography at ‘C’ as the angle of illumination is just above
grazing incidence. Consequently the slightest change in surface can cause
a significant shadow to be cast (analogous to long shadows being cast at
sunset). As the region that is being observed gets further above the centre
line, ‘C’, the angle of illumination becomes more normal to the ‘local’
plane of the surface and the shadowing becomes less pronounced (small shadows
cast at noon) and the measurements are then less sensitive to the surface
topography.
The image acquired by the camera is stored as a two-dimensional array
of pixel intensities. Along each array, row and column, there is a slowly
varying component of intensity caused by the illumination geometry, as
well as higher frequency components caused by roughness. The slowly varying
component is removed.
The roughness indices are based on the standard deviation or RMS variation.
The RMS variation in intensity is first calculated as
xRMS =[ (1/N) Σxi2 ] 1/2
where xi represents the difference between the intensity value of each
pixel, i, and the average corrected intensity for a particular array. The
individual array RMS values are weighted and then averaged to give an average
‘roughness index’.
TISSUE APPLICATION
Sanitary tissue papers cannot be measured using common air-leak roughness
methods. These grades are too porous and/or compressible. A one-ply tissue
typically has a grammage below 15 g/m2 and lacks the stiffness to be fed
into the apparatus while being held snugly against the backing cylinder
(Figure 1). A method was devised where the tissue was mounted on a flexible
flat backing sheet with a slight in-plane tension. The mounted tissue was
then inserted into the OpTiSurf for analysis.
Facial tissue softness: A Canadian tissue manufacturer provided three
sets of two-ply facial tissue. A panel of judges at the manufacturer had
ranked these as soft, acceptable and roughest. There is an expectation
that surface roughness is one of the attributes that impacts softness.
The sheets were separated into single plies and the top ply mounted for
testing. The MD direction of the tissue was tested as to ensure that the
crepe was included in the analysis. The optical roughness index (ORI) increased
as the panel ranking indicated that sheet was less soft.
The OpTiSurf provided FFT intensity spectra data in the form of seven
Roughness Intensity components but the relative roughness is most interesting
to study (Figure 2). This demonstrates how the roughness of the ‘softest’
and ‘roughest’ sample relative to the ‘acceptable’ sheet changes depending
the size of the roughness component. At about 12 mm, corresponding to OpTiSurf
component R6, the ranking is comparable to the panel ranking of softness.
This may indicate that the judges of the softness panel are sensitive to
the roughness at a 8 mm to 16 mm scale.
Bathroom tissue embossing: A bathroom tissue manufacturer required a
quantifiable way of determining the variation of embossing intensity. It
was anticipated that the relative roughness would serve as a measurable
indicator of embossing intensity. The intensity of the embossing may vary
from the inner layers of the bathroom tissue (BRT) roll, near the core,
to the outer layers. Also, the embossing quality may vary from roll to
roll. Two sets of commercial BRT rolls were provided. Each set contained
four BRT rolls. One set had acceptable embossing while the other did not.
From each roll a sheet was sampled at: three layers from the core, 1/4”
from the core, 2/3” from the core and three layers from the outer layer.
The sheet were mounted and measured both in the MD and CD.
Figure 3 reveals that the ORI, and thus the embossing intensity, diminish
when going from the BRT core to the outer ply. Also the ORI was greater
for the acceptable samples (triangle identifiers in Figure 3) compared
to the unacceptable samples.
The MD and CD roughness Intensity spectra (Figures 4 & 5), reveal
the greatest difference between the acceptable and unacceptable embossing
intensity occurred at the R4 size component (2-4 mm) and likely corresponds
to the embossing dimensions. The relative roughness intensities at size
component R4 were calculated at each BRT roll diameter position (Figure
6). The acceptable sample was used as the reference sheet. A value less
than 1 indicates that the test sheet has a lower roughness (embossing)
intensity relative to the reference sheet.
Figure 6 shows that, on average, the R4 MD relative embossing intensity
was about 60% of the acceptable level and that the CD relative embossing
intensity was about 45% of the acceptable level. This approach provides
a bathroom tissue manufacture with a quantitative method of monitoring
embossing intensity.